PXB Error Messages  

There are two types of error messages that you can encounter.

Introduction

If an error condition occurs in the PXB, it is reported to both the front panel display error queue and the SCPI (remote interface) error queue. These two queues are viewed and managed separately.

Characteristic

Front Panel Display
Error Queue

SCPI Remote Interface
Error Queue

Capacity (#errors)

30

30

Overflow Handling

Circular (rotating).
Drops oldest error as new error comes in.

Linear, first in/first out.

Viewing Entries

Select the information button in the lower left corner of the user interface.

Use SCPI query
SYSTem:ERRor[:NEXT]?

Clearing the Queue

Select: Tools > Clear Error Queue

Power up
Send a
*CLS command
Read last item in the queue.

Unresolved Errors
(errors that must be resolved.)

Re-reported after queue is cleared.

Re-reported after queue is cleared. 

 

PXB Error Messages

These error messages are displayed on the PXB during operation.

Number Name/ID Message Comment

-222

Data Out Of Range

{0} clipped to lower limit.

{0}=Setting Name

-222

Data Out Of Range

{0} clipped to upper limit.

{0}=Setting Name

-222

Data Out Of Range

Baseband Generator: {0} clipped to upper limit.

{0}=Setting Name

-222

Data Out Of Range

Correlation magnitude was clipped to the maximum value of 1.

 

-222

Data Out Of Range

DSIM: {0} clipped to lower limit.

{0}=Setting Name

-222

Data Out Of Range

DSIM: {0} clipped to upper limit.

{0}=Setting Name

-222

Data Out Of Range

I/O: {0} clipped to lower limit.

{0}=Setting Name

-222

Data Out Of Range

I/O: {0} clipped to upper limit.

{0}=Setting Name

-222

Data Out Of Range

Signal Analyzer: {0} clipped to lower limit.

{0}=Setting Name

-222

Data Out Of Range

Signal Analyzer: {0} clipped to upper limit.

{0}=Setting Name

-222

Data Out Of Range

Signal Generator: {0} clipped to lower limit.

{0}=Setting Name

-222

Data Out Of Range

Signal Generator: {0} clipped to upper limit.

{0}=Setting Name

-222

Data Out Of Range

The value {0} was clipped to the minimum value {1}

{0}=Setting value
{1}=Min value

-221

Settings Conflict

Baseband Generator: Cannot modify 'non runtime' setting during power calibration.

 

-221

Settings Conflict

Baseband Generator: Cannot modify 'non runtime' setting while playing.

 

-221

Settings Conflict

Baseband Generator: Cannot modify disabled setting.

 

-221

Settings Conflict

Baseband Generator: Cannot modify or query settings. Instrument not configured or Baseband Generator not enabled.

 

-221

Settings Conflict

Baseband Generator: Cannot modify Power Meter settings in Auto Power Calibration mode.

 

-221

Settings Conflict

Baseband Generator: Cannot modify read-only setting.

 

-221

Settings Conflict

Baseband Generator: Cannot modify setting. Marker 2 is reserved for internal use.

 

-221

Settings Conflict

Baseband Generator: Cannot modify setting. Waveform created by Signal Studio.

 

-221

Settings Conflict

Baseband Generator: Cannot modify waveform or marker files while downloading.

 

-221

Settings Conflict

Cannot modify clock settings. The instrument is busy.

 

-221

Settings Conflict

Cannot modify input setting while DSIM channel is configured as output.

 

-221

Settings Conflict

Cannot modify output setting while DSIM channel is configured as input.

 

-221

Settings Conflict

Cannot modify trigger polarity when trigger source is not HW trigger or when the instrument is busy.

 

-221

Settings Conflict

Cannot modify trigger settings. The instrument is busy.

 

-221

Settings Conflict

DSIM: Cannot modify 'non runtime' setting during power calibration.

 

-221

Settings Conflict

DSIM: Cannot modify 'non runtime' setting while playing.

 

-221

Settings Conflict

DSIM: Cannot modify disabled setting.

 

-221

Settings Conflict

DSIM: Cannot modify or query settings. Instrument not configured or DSIM not enabled.

 

-221

Settings Conflict

DSIM: Cannot modify Parallel mode setting when in Serial mode.

 

-221

Settings Conflict

DSIM: Cannot modify read only setting.

 

-221

Settings Conflict

DSIM: Cannot modify reference frequency source. The setting is read only when clock source is set to external.

 

-221

Settings Conflict

DSIM: Cannot modify reference frequency. The setting is read only at this time.

 

-221

Settings Conflict

DSIM: Cannot modify Serial mode setting when in Parallel mode.

 

-221

Settings Conflict

DSIM: Cannot modify Word Alignment. The setting is disabled when word size is 16 bits or larger.

 

-221

Settings Conflict

Fading: Antenna settings are read only unless the correlation source is Custom Antenna Setup.

 

-221

Settings Conflict

Fading: Beam forming correlation type is only valid for 4x2 and 2x4 configurations.

 

-221

Settings Conflict

Fading: Cannot modify 'non runtime' setting while playing.

 

-221

Settings Conflict

Fading: Cannot modify disabled channel.

 

-221

Settings Conflict

Fading: Cannot modify disabled path.

 

-221

Settings Conflict

Fading: Cannot modify disabled setting.

 

-221

Settings Conflict

Fading: Cannot modify or query settings. Instrument not configured or Fading not enabled.

 

-221

Settings Conflict

Fading: Cannot modify read only setting.

 

-221

Settings Conflict

Fading: Correlation coefficients are read only unless the correlation source is Custom Correlation Matrix.

 

-221

Settings Conflict

Fading: Downlink direction is invalid for 2x4 configuration with Beam forming.

 

-221

Settings Conflict

Fading: High, Medium, and Low correlation types are only valid for 2x2 configuration.

 

-221

Settings Conflict

Fading: Only Laplacian Power Azimuth Spectrum is allowed for High, Medium, or Low correlation.

 

-221

Settings Conflict

Fading: Path Configuration settings are read only unless the path configuration source is Custom.

 

-221

Settings Conflict

Fading: The correlation type is read only unless the correlation source is set to Standard.

 

-221

Settings Conflict

Fading: The first channel index must be larger than the second channel index when setting a correlation coefficient.

 

-221

Settings Conflict

Fading: Uplink direction is invalid for 4x2 configuration with Beam forming.

 

-221

Settings Conflict

I/O: Cannot modify 'non runtime' setting during power calibration.

 

-221

Settings Conflict

I/O: Cannot modify 'non runtime' setting while playing.

 

-221

Settings Conflict

I/O: Cannot modify Analog Enabled when configured as a Digital Port.

 

-221

Settings Conflict

I/O: Cannot modify AWGN Settings. AWGN not licensed.

 

-221

Settings Conflict

I/O: Cannot modify input settings while I/O channel is configured as Output.

 

-221

Settings Conflict

I/O: Cannot modify LVDS Enabled when configured as an Analog Port.

 

-221

Settings Conflict

I/O: Cannot modify Marker 4 Settings when connected to a DSIM. The DSIM only uses Markers 1 and 3.

 

-221

Settings Conflict

I/O: Cannot modify or query settings. The instrument not configured or I/O not enabled.

 

-221

Settings Conflict

I/O: Cannot modify output settings while I/O channel is configured as Input.

 

-221

Settings Conflict

I/O: Cannot modify Power Meter settings in Auto Power Calibration mode.

 

-221

Settings Conflict

I/O: Cannot modify read only setting.

 

-221

Settings Conflict

I/O: Cannot modify setting. Marker 2 is reserved for internal use.

 

-221

Settings Conflict

Power meter: Setting disabled when Input RMS Source is not set to Measure.

 

-221

Settings Conflict

Power meter: Setting disabled when Input RMS Source is not set to User.

 

-221

Settings Conflict

Power meter: Setting disabled when Power Meter Gate is set to Threshold.

 

-221

Settings Conflict

Power meter: Setting disabled when Power Meter Gating is set to Marker.

 

-221

Settings Conflict

Power meter: Setting is read only when Input RMS Source is set to Waveform.

 

-221

Settings Conflict

Signal Analyzer: Cannot modify 'non runtime' setting during power calibration.

 

-221

Settings Conflict

Signal Analyzer: Cannot modify 'non runtime' setting while playing.

 

-221

Settings Conflict

Signal Analyzer: Cannot modify disabled setting.

 

-221

Settings Conflict

Signal Analyzer: Cannot modify or query settings. Instrument not configured or Signal Analyzer not enabled.

 

-221

Settings Conflict

Signal Analyzer: Cannot modify read only setting.

 

-221

Settings Conflict

Signal Generator: Cannot modify 'non runtime' setting during power calibration.

 

-221

Settings Conflict

Signal Generator: Cannot modify 'non runtime' setting while playing.

 

-221

Settings Conflict

Signal Generator: Cannot modify disabled setting.

 

-221

Settings Conflict

Signal Generator: Cannot modify or query settings. Instrument not configured or Signal Generator not enabled.

 

-221

Settings Conflict

Signal Generator: Cannot modify read only setting.

 

-221

Settings Conflict

The instrument is busy.

 

-150

String Data Error

Invalid master channel name: {0}.

{0}=Master Channel Name

1

Startup Error

Interconnect board not found.

 

1

Startup Error

Invalid BB board arrangement. Arrange the boards to be contiguous beginning with slot 1 (back, next to clock board).

Message is displayed when there is a non contiguous Baseband board in the PXB system.

1

Startup Error

Invalid I/O slot number: {0}.

{0}=Slot number

1

Startup Error

One or more EEPROM boards not programmed correctly. If the problem persists, contact Agilent Technologies.

 

1

Startup Error

Power up self test failed. Run full self test for more details.

 

1

Startup Error

System hardware failure. Power down and restart the instrument. If the problem persists, contact Agilent Technologies.

 

2

Shutdown Error

An unexpected error has occurred. Power down and restart the instrument. If the problem persists, contact Agilent Technologies.

 

3

Self Test Error

Factory self test data not available. Power down and restart the instrument. If the problem persists, contact Agilent Technologies.

 

3

Self Test Error

Failed to run test {0} of test group {1}. Power down and restart the instrument. If the problem persists, contact Agilent Technologies.

{0}=Test Name
{1}=Test Group

3

Self Test Error

Failed to save test results. Power down and restart the instrument. If the problem persists, contact Agilent Technologies.

 

4

Play Error

Configuration license expired.

 

4

Play Error

Hardware optimization failed.

 

4

Play Error

Invalid configuration.

 

4

Play Error

Load Baseband Generator Setting Error. System hardware failure. Power down and restart the instrument. If the problem persists, contact Agilent Technologies.

 

4

Play Error

Load DSIM Setting Error. System hardware failure. Power down and restart the instrument. If the problem persists, contact Agilent Technologies.

 

4

Play Error

Load Fading Setting Error. Could not calculate the Cholesky decomposition for the correlation matrix.

 

4

Play Error

Load Fading Setting Error. Fading model not licensed.

 

4

Play Error

Load Fading Setting Error. Invalid dynamic fading file.

 

4

Play Error

Load Fading Setting Error. System hardware failure. Power down and restart the instrument. If the problem persists, contact Agilent Technologies.

 

4

Play Error

Load I/O Setting Error. System hardware failure. Power down and restart the instrument. If the problem persists, contact Agilent Technologies.

 

4

Play Error

Load Signal Analyzer Setting Error. External instrument {0} not connected or not powered on.

{0}=Instrument Name

4

Play Error

Load Signal Generator Setting Error. External instrument {0} not connected or not powered on.

{0}=Instrument Name

4

Play Error

Signal generator {0} did not arm properly. Check digital cable between PXB and signal generator {0}.

{0}=Instrument Name

4

Play Error

Signal generator {0} did not arm in the allowed amount of time.

{0}=Instrument Name

4

Play Error

The instrument not configured yet.

 

5

Stop Error

An unexpected error has occurred. Power down and restart the instrument. If the problem persists, contact Agilent Technologies.

 

6

Save Error

Could not open file {0}.

{0}=File Name

6

Save Error

Save Baseband Generator Setting Error. Power down and restart the instrument. If the problem persists, contact Agilent Technologies.

 

6

Save Error

Save DSIM Setting Error. Power down and restart the instrument. If the problem persists, contact Agilent Technologies.

 

6

Save Error

Save Fading Setting Error. Power down and restart the instrument. If the problem persists, contact Agilent Technologies.

 

6

Save Error

Save I/O Setting Error. Power down and restart the instrument. If the problem persists, contact Agilent Technologies.

 

6

Save Error

Save Signal Analyzer Setting Error. Power down and restart the instrument. If the problem persists, contact Agilent Technologies.

 

6

Save Error

Save Signal Generator Setting Error. Power down and restart the instrument. If the problem persists, contact Agilent Technologies.

 

7

Recall Error

Could not open file {0}.

{0}=File Name

7

Recall Error

Recall Baseband Generator Setting Error. Saved state corrupted or out of date.

 

7

Recall Error

Recall Configuration Error. Saved state corrupted or out of date.

 

7

Recall Error

Recall DSIM Setting Error. Saved state corrupted or out of date.

 

7

Recall Error

Recall Fading Setting Error. Could not calculate the Cholesky decomposition for the correlation matrix.

 

7

Recall Error

Recall Fading Setting Error. Fading model not licensed.

 

7

Recall Error

Recall Fading Setting Error. Invalid dynamic fading file.

 

7

Recall Error

Recall Fading Setting Error. Saved state corrupted or out of date.

 

7

Recall Error

Recall I/O Setting Error. Saved state corrupted or out of date.

 

7

Recall Error

Recall Signal Analyzer Setting Error. Saved state corrupted or out of date.

 

7

Recall Error

Recall Signal Generator Setting Error. Saved state corrupted or out of date.

 

8

Copy Error

Copy Baseband Generator Setting Error.

 

8

Copy Error

Copy DSIM Setting Error.

 

8

Copy Error

Copy Fading Setting Error.

 

8

Copy Error

Copy I/O Setting Error.

 

8

Copy Error

Copy Signal Analyzer Setting Error.

 

8

Copy Error

Copy Signal Generator Setting Error.

 

9

Paste Error

Paste Baseband Generator Setting Error.

 

9

Paste Error

Paste DSIM Setting Error.

 

9

Paste Error

Paste Fading Setting Error.

 

9

Paste Error

Paste I/O Setting Error.

 

9

Paste Error

Paste Signal Analyzer Setting Error.

 

9

Paste Error

Paste Signal Generator Setting Error.

 

10

Reset Error

An unexpected error has occurred. Power down and restart the instrument. If the problem persists, contact Agilent Technologies.

 

20

Select Configuration Error

Configuration corrupted.

 

20

Select Configuration Error

Configuration not found.

 

20

Select Configuration Error

Failed to retrieve installed licenses.

 

20

Select Configuration Error

Invalid configuration.

 

21

Load Configuration Error

Cannot apply config 'FADE (ext in) 1 Channel' because it has conflicting instrument assignments.

Possible licensing problem (see Loopback Fading)

21

Load Configuration Error

Cannot use GPIB interface {0} as a system controller. It is not configured as a system controller.

 

21

Load Configuration Error

Configuration not licensed.

 

21

Load Configuration Error

DSIM not initialized properly. An unexpected error has occurred. Power down and restart the instrument. If the problem persists, contact Agilent Technologies.

 

21

Load Configuration Error

External instrument {0} and {1} have the same address ({2}).

{0}=Instrument1 Name
{1}=Instrument2 Name
{2}=Instrument Address

21

Load Configuration Error

Failed to apply Fading model. Fading model {0} not licensed.

{0}=Model Name

21

Load Configuration Error

Failed to apply Fading model. Failed to calculate the correlation coefficients.

 

21

Load Configuration Error

Failed to retrieve installed licenses.

 

21

Load Configuration Error

FPGA Alignment Failure

 

22

Modify Configuration Error

Analog input port not supported.

 

22

Modify Configuration Error

Configuration not found.

 

22

Modify Configuration Error

GPIB card {0} is configured as controller.

{0} GPIB card number

22

Modify Configuration Error

GPIB card not found.

 

22

Modify Configuration Error

Instrument {0} not found.

{0}=Instrument Name

22

Modify Configuration Error

Invalid analog output instrument type. Signal Generator or Custom instrument expected.

 

22

Modify Configuration Error

Invalid digital input instrument type. DSIM or Signal Analyzer expected.

 

22

Modify Configuration Error

Invalid digital output instrument type. DSIM or Signal Generator expected.

 

22

Modify Configuration Error

I/O port {0} not available.

{0}=I/O Port Name

22

Modify Configuration Error

I/O port {0} not found.

{0}=I/O Port Name

30

Download Waveform Error

Cannot start a download when another download is in progress.

 

30

Download Waveform Error

Could not open file {0}.

{0}=Waveform File Name

30

Download Waveform Error

Failed to transfer waveform and marker data from files to dram.

 

30

Download Waveform Error

Marker file not found: {0}.

{0}=Marker File Name

30

Download Waveform Error

The number of samples and the number of markers does not match.

 

30

Download Waveform Error

Waveform not found: {0}.

{0}=Waveform File Name

30

Download Waveform Error

Waveform not licensed: {0}.

{0}=Waveform Name

31

Transfer Data Error

Failed to transfer waveform data to instrument. Waveform name: {0}.

{0}=Waveform Name

40

Power Calibration Error

Failed to abort power calibration.

 

40

Power Calibration Error

Failed to restore system state after a power calibration error.

 

40

Power Calibration Error

Instrument not configured yet.

 

40

Power Calibration Error

Power Calibration cannot complete, all inputs are disabled.

 

40

Power Calibration Error

Power Calibration setup failed for signal generator {0}.

{0}=Instrument Name

40

Power Calibration Error

Power Meter Gating is set to Marker and the marker is turned off.

 

40

Power Calibration Error, Power Search Error

Power search failed for signal generator {0}. Increase I/Q signal level or reduce signal generator amplitude.

{0}=Instrument Name

40

Power Calibration Error

Reading Power Meters timed out.

 

40

Power Calibration Error

The instrument is busy.

 

40

Power Calibration Error

Waveform RMS level exceeds signal generator {0} power search manual level maximum of {1}.

{0}=Instrument Name
{1}=Max RMS

40

Power Calibration Error

Waveform RMS level is too low for successful signal generator {0} power search.

{0}=Instrument Name

41

I/O Calibration Error

Failed to restore factory calibration data. An unexpected error has occurred. Power down and restart the instrument. If the problem persists, contact Agilent Technologies.

 

41

I/O Calibration Error

The instrument is busy.

 

50

External Reference Input Error

External reference frequency out of range.

 

50

External Reference Input Error

External reference source not detected.

 

60

External Trigger Input Error

Report this error to Agilent Technologies.

 

100

System Error

External reference source not detected. PLL unlocked.

 

100

System Error

Failed to configure GPIB address. GPIB card not found.

 

100

System Error

Failed to toggle touch screen on/off. Touch screen was not installed properly.

 

100

System Error

Invalid external reference frequency. PLL unlocked.

 

100

System Error

PLL unlocked.

 

101

Hardware Error

Clock board temperature {0} exceeds threshold. Shutting down system.

{0}=Temperature

101

Hardware Error

Fans have stopped working. Shutting down system.

 

101

Hardware Error

I/O board temperature {0} exceeds threshold. Shutting down system.

{0}=Temperature

101

Hardware Error

Mother board temperature {0} exceeds threshold. Shutting down system.

{0}=Temperature

101

Hardware Error

Power supply temperature {0} exceeds threshold. Shutting down system.

{0}=Temperature

102

Add External Instrument Error

Instrument {0} already exists.

{0}=Instrument Name

102

Add External Instrument Error

Invalid instrument name.

{0}=Instrument Name

102

Add External Instrument Error

Signal analyzer list is full. Remove an existing signal analyzer before adding a new one.

 

102

Add External Instrument Error

Signal generator list is full. Remove an existing signal generator before adding a new one.

 

102

Add External Instrument Error

The custom instrument list is full. Remove an existing custom instrument before adding a new one.

 

102

Add External Instrument Error

The DSIM list is full. Remove an existing DSIM before adding a new one.

 

103

Modify External Instrument Error

Instrument {0} not found.

{0}=Instrument Name

104

Remove External Instrument Error

Instrument {0} in use.

{0}=Instrument Name

104

Remove External Instrument Error

Instrument {0} not found.

{0}=Instrument Name

105

Test Connection Error

Communication failure with instrument {0}. Instrument is busy, not connected or not powered on.

{0}=Instrument Name

110

Launch Application Error

Failed to launch {0}.

{0}=Application Name

111

Launch License Manager Error

Agilent License Server not installed properly.

 

111

Launch License Manager Error

Agilent License Server not installed.

 

200

Warning

{0} does not have external instrument yet. Please assign an instrument to this I/O port.

{0}=I/O Port Name

200

Warning

Address {0} is already assigned to external instrument {1}.

{0}=Instrument Address
{1}=Instrument Name

200

Warning

Backup file corrupted or out of date.

 

200

Warning

Correlation magnitude was clipped to the maximum value of 1.

 

200

Warning

Correlation value must be a complex number with magnitude less than or equal to 1.

 

200

Warning

External instrument {0} and {1} have the same address ({2}).

{0}=Instrument1 Name
{1}=Instrument2 Name
{2}=Instrument Address

200

Warning

Input Frequency Offset + Input Signal Bandwidth may exceed the effective fader bandwidth. Reduce the frequency offset or number of enabled fader paths. 

 

200

Warning

Input Frequency Offset + Input Signal Bandwidth may exceed the output device bandwidth.

 

200

Warning

Input Frequency Offset + Output Frequency Offset may exceed the output device bandwidth.

 

200

Warning

Input Frequency Offset exceeds the effective fader bandwidth. Reduce the frequency offset or decrease the number of enabled fader paths.

 

200

Warning

Instrument {0} has invalid address. Correct at a later time.

{0}=Instrument Name

200

Warning

Instrument {0} not connected or not powered on.

{0}=Instrument Name

200

Warning

LXI LAN Reset Error

 

200

Warning

No hardware or license found.

 

200

Warning

No markers active. If you intend to use markers for ALC or RF blanking, you must turn them on.

 

200

Warning

Output Frequency Offset + Noise Bandwidth exceeds the output device bandwidth.

 

200

Warning

Power Meter Gating is set to Marker and the marker is turned off.

 

200

Warning

Recall external instruments failed.

 

200

Warning

The AWGN noise bandwidth exceeds the output device bandwidth. Reduce the noise bandwidth to {0} MHz.

{0}=Output Device Bandwidth

200

Warning

The input frequency offset exceeds the output device bandwidth. Reduce the input frequency offset.

 

200

Warning

Touch screen not installed properly.

 

200

Warning

Waveform not licensed. Waveform name: {0}.

{0}=Waveform Name

201

Information

{0} was clipped to {1} due to the valid resolution changing.

{0}=Setting Name
{1}=Setting Value

201

Information

{0} was clipped to the maximum value of {1}

{0}=Setting Name
{1}=Setting Value

201

Information

{0} was clipped to the minimum value of {1}

{0}=Setting Name
{1}=Setting Value

201

Information

Baseband Generator: {0} clipped to {1}.

{0}=Setting Name
{1}=Setting Value

201

Information

DSIM: {0} clipped to {1}.

{0}=Setting Name
{1}=Setting Value

201

Information

I/O: Failed to copy master channel list.

 

201

Information

I/O: {0} clipped to {1}.

{0}=Setting Name
{1}=Setting Value

201

Information

Power Calibration will run automatically on Play.

 

201

Information

Power control settings have been changed to match the current waveform type.

 

201

Information

Signal analyzer {0} is connected.

{0}=Instrument Name

201

Information

Signal analyzer {0} is not connected. Address not found.

{0}=Instrument Name

201

Information

Signal analyzer {0} is not connected. Address not valid.

{0}=Instrument Name

201

Information

Signal analyzer {0} is not connected. Connection lost.

{0}=Instrument Name

201

Information

Signal analyzer {0} is not connected. Model number not supported.

{0}=Instrument Name

201

Information

Signal analyzer {0} is not connected. Not a member of the {0} family.

{0}=Instrument Family

201

Information

Signal analyzer {0} is not connected. Requires firmware version {0} or higher and option (one of) {1}.

{0}=Firmware version
{1}=Option

201

Information

Signal analyzer {0} is not connected. Requires firmware version {0} or higher.

{0}=Firmware version

201

Information

Signal analyzer: {0} clipped to {1}.

{0}=Setting Name
{1}=Setting Value

201

Information

Signal generator {0} is connected.

{0}=Instrument Name

201

Information

Signal generator {0} is not connected. Address not found.

{0}=Instrument Name

201

Information

Signal generator {0} is not connected. Address not valid.

{0}=Instrument Name

201

Information

Signal generator {0} is not connected. Connection lost.

{0}=Instrument Name

201

Information

Signal generator {0} is not connected. Model number not supported.

{0}=Instrument Name

201

Information

Signal generator {0} is not connected. Not a member of the {0} family.

{0}=Instrument Family

201

Information

Signal generator {0} is not connected. Requires firmware version {0} or higher and option (one of) {1}.

{0}=Firmware version
{1}=Option

201

Information

Signal generator {0} is not connected. Requires firmware version {0} or higher.

{0}=Firmware version

201

Information

Signal generator: {0} clipped to {1}.

{0}=Setting Name
{1}=Setting Value

201

Information

Signal power calculations are invalid until a new power calibration is run.

 

300

Modify Baseband Generator Setting Error

Cannot modify waveform source created by Signal Studio.

 

300

Modify Baseband Generator Setting Error

I/P RMS Source invalid. Cannot extract I/P RMS information from binary waveforms.

 

300

Modify Baseband Generator Setting Error

Invalid marker file name: {0}.

{0}=Marker File Name

300

Modify Baseband Generator Setting Error

Invalid waveform file name: {0}.

{0}=Waveform File Name

300

Modify Baseband Generator Setting Error

Marker file not found: {0}.

{0}=Marker File Name

300

Modify Baseband Generator Setting Error

Waveform file not found: {0}.

{0}=Waveform File Name

300

Modify Baseband Generator Setting Error

Waveform size too small. It must have at least 256 samples.

 

301

Load Baseband Generator Setting Error

System hardware failure.

 

302

Restore Baseband Generator Default Setting Error

 

 

400

Modify Fading Setting Error

Birth Death disabled when dynamic fading is enabled.

 

400

Modify Fading Setting Error

Dynamic Fading not allowed for faders using Birth Death.

 

400

Modify Fading Setting Error

Dynamic Fading not allowed for faders using High Speed Train.

 

400

Modify Fading Setting Error

Dynamic Fading not allowed for faders using Moving Propagation.

 

400

Modify Fading Setting Error

Fading model not licensed: {0}.

{0}=Model Name

400

Modify Fading Setting Error

High bits of Random Seed must be between 0x0 and 0x1FFFFFF.

 

400

Modify Fading Setting Error

High Speed Train disabled when dynamic fading is enabled.

 

400

Modify Fading Setting Error

Low bits of Random Seed must be between 0x0 and 0xFFFFFFFF.

 

400

Modify Fading Setting Error

Middle bits of Random Seed must be between 0x0 and 0xFFFFFFFF.

 

400

Modify Fading Setting Error

Moving Propagation disabled when dynamic fading is enabled.

 

401

Load Fading Setting Error

Delay out of range. Expected [{0}, {1}]; found {2}

{0} = minimum delay (currently 0 sec)
{1} = maximum delay (currently 2 ms)
{2} = delay found

401

Load Fading Setting Error

Dwell Time out of range. Expected [{0}, {1}]; found {2}

{0} = minimum dwell time (currently 0.01 sec)
{1} = maximum dwell time (currently 1000 sec)
{2} = dwell time found

401

Load Fading Setting Error

Fading Settings Download already in progress.

 

401

Load Fading Setting Error

Failed to calculate the Cholesky decomposition for the correlation matrix.

 

401

Load Fading Setting Error

Invalid dynamic fading file.

 

401

Load Fading Setting Error

Invalid dynamic fading file. Expected {0} parameters on line {1} ({2} columns x {3} paths), but {4} were found.  

{0} = expected number of parameters
{1} = line number
{2} = number of columns
{3} = number of paths
{4} = number of parameters found

401

Load Fading Setting Error

Invalid dynamic fading file. Number of dynamic fading states exceeds maximum of {0}. 

{0} = Maximum number of dynamic fading states allowed (currently 5000)

401

Load Fading Setting Error

Invalid Dynamic Fading File. The vehicle speed for each path must be the same unless the spectral shape is Jakes Rounded or Jakes Classical. 

 

401

Load Fading Setting Error

Invalid Fader directive in dynamic fading file on line {0} - couldn't read number of paths.

{0} = line number

401

Load Fading Setting Error

Loss out of range. Expected [{0}, {1}]; found {2} 

{0} = minimum loss (currently 0 dB)
{1} = maximum loss (currently 84 dB)
{2} = loss found

401

Load Fading Setting Error

System hardware failure.

 

401

Load Fading Setting Error

UE Speed out of range. Expected [{0}, {1}]; found {2} 

{0} = minimum UE speed (currently 0 km/hr)
{1} = maximum delay (based on the Doppler frequency and carrier frequency)
{2} = UE speed found

401

Load Fading Setting Error

Undefined parameter name in Columns directive of dynamic fading file on line {0}.

{0} = line number

402

Restore Fading Default Setting Error

 

 

403

Apply Fading Model Error

Fading model not licensed: {0}.

{0}=Model Name

403

Apply Fading Model Error

Failed to calculate the correlation coefficients.

 

500

Modify I/O Setting Error

Cannot modify input setting while I/O channel is configured as output.

 

500

Modify I/O Setting Error

Cannot modify output setting while I/O channel is configured as input.

 

500

Modify I/O Setting Error

Failed to set {0} to {1} Setting invalid. Signal power calculations are invalid until a new power calibration is run.

{0}=Setting Name
{1}=Setting Value

501

Load I/O Setting Error

System hardware failure.

 

502

Restore I/O Default Setting Error

 

 

601

Load Signal Generator Setting Error

{0} not connected or not powered on.

{0}=Instrument Name

607

Query Signal Generator Setting Error

{0} not connected or not powered on.

{0}=Instrument Name

608

Signal Generator Power Search Error

{0} not connected or not powered on.

{0}=Instrument Name

609

Preset Signal Generator Error

{0} not connected or not powered on.

{0}=Instrument Name

701

Load Signal Analyzer Setting Error

{0} not connected or not powered on.

{0}=Signal Analyzer Name

702

Preset Signal Analyzer Error

{0} not connected or not powered on.

{0}=Signal Analyzer Name

703

Query Signal Analyzer Setting Error

{0} not connected or not powered on.

{0}=Signal Analyzer Name

801

Load DSIM Setting Error

DSIM not connected or not powered on.

 

801

Load DSIM Setting Error

System hardware failure.

 

802

Restore DSIM Default Setting Error

DSIM not connected or not powered on.

 

802

Restore DSIM Default Setting Error

System hardware failure.

 

 

Dynamic Fading Excel Spreadsheet Error Messages

These error messages can be displayed when you are completing the Microsoft Excel spreadsheet template for dynamic fading.

Message Comment

Based on the carrier frequency of {0} and the maximum Doppler frequency of 1600 Hz, the UE Speed for Path {1} in row {2} must be between 0 and {3} km/hr.

{0} = carrier frequency
{1} = path number
{2} = row number
{3} = maximum UE speed

Delay for Path {0} in row {1} must be a numeric value.

{0} = path number
{1} = row number

Delay for Path {0} in row {1} must be between 0 s and 2 ms.

{0} = path number
{1} = row number

Dwell Time for row {0} must be a numeric value.

{0} = row number

Dwell Time for row {0} must be greater than or equal to 0.01 s.

{0} = row number

Error encountered during Export. Please correct the error and try again. 

 

Fader {0} Carrier Frequency must be a numeric value.

{0} = fader number

Loss for Path {0} in row {1} must be a numeric value.

{0} = path number
{1} = row number

Loss for Path {0} in row {1} must be between 0 and 84 dB.

{0} = path number
{1} = row number 

The file you have specified for export is marked as read only.

 

The maximum number of dynamic fading states 5000, but fader {0} has {1} states. Exporting the first 5000 states. 

{0} = fader number
{1} = number of states found

UE Speed for Path {0} in row {1} must be a numeric value. 

{0} = path number
{1} = row number