There are two types of error messages that you can encounter.
PXB Error Messages are displayed on the PXB during operation to indicate that there is a problem.
Dynamic Fading
Excel Spreadsheet Error Messages are displayed on the computer being
used to enter the dynamic fading data into the dynamic fading Microsoft
Excel spreadsheet.
If an error condition occurs in the PXB, it is reported to both the front panel display error queue and the SCPI (remote interface) error queue. These two queues are viewed and managed separately.
Characteristic |
Front Panel Display
|
SCPI Remote Interface
|
Capacity (#errors) |
30 |
30 |
Overflow Handling |
Circular (rotating). |
Linear, first in/first out. |
Viewing Entries |
Select the information button in the lower left corner of the user interface. |
Use SCPI query |
Clearing the Queue |
Select: Tools > Clear Error Queue |
Power up |
Unresolved Errors |
Re-reported after queue is cleared. |
Re-reported after queue is cleared. |
These error messages are displayed on the PXB during operation.
Number | Name/ID | Message | Comment |
---|---|---|---|
-222 |
Data Out Of Range |
{0} clipped to lower limit. |
{0}=Setting Name |
-222 |
Data Out Of Range |
{0} clipped to upper limit. |
{0}=Setting Name |
-222 |
Data Out Of Range |
Baseband Generator: {0} clipped to upper limit. |
{0}=Setting Name |
-222 |
Data Out Of Range |
Correlation magnitude was clipped to the maximum value of 1. |
|
-222 |
Data Out Of Range |
DSIM: {0} clipped to lower limit. |
{0}=Setting Name |
-222 |
Data Out Of Range |
DSIM: {0} clipped to upper limit. |
{0}=Setting Name |
-222 |
Data Out Of Range |
I/O: {0} clipped to lower limit. |
{0}=Setting Name |
-222 |
Data Out Of Range |
I/O: {0} clipped to upper limit. |
{0}=Setting Name |
-222 |
Data Out Of Range |
Signal Analyzer: {0} clipped to lower limit. |
{0}=Setting Name |
-222 |
Data Out Of Range |
Signal Analyzer: {0} clipped to upper limit. |
{0}=Setting Name |
-222 |
Data Out Of Range |
Signal Generator: {0} clipped to lower limit. |
{0}=Setting Name |
-222 |
Data Out Of Range |
Signal Generator: {0} clipped to upper limit. |
{0}=Setting Name |
-222 |
Data Out Of Range |
The value {0} was clipped to the minimum value {1} |
{0}=Setting value |
-221 |
Settings Conflict |
Baseband Generator: Cannot modify 'non runtime' setting during power calibration. |
|
-221 |
Settings Conflict |
Baseband Generator: Cannot modify 'non runtime' setting while playing. |
|
-221 |
Settings Conflict |
Baseband Generator: Cannot modify disabled setting. |
|
-221 |
Settings Conflict |
Baseband Generator: Cannot modify or query settings. Instrument not configured or Baseband Generator not enabled. |
|
-221 |
Settings Conflict |
Baseband Generator: Cannot modify Power Meter settings in Auto Power Calibration mode. |
|
-221 |
Settings Conflict |
Baseband Generator: Cannot modify read-only setting. |
|
-221 |
Settings Conflict |
Baseband Generator: Cannot modify setting. Marker 2 is reserved for internal use. |
|
-221 |
Settings Conflict |
Baseband Generator: Cannot modify setting. Waveform created by Signal Studio. |
|
-221 |
Settings Conflict |
Baseband Generator: Cannot modify waveform or marker files while downloading. |
|
-221 |
Settings Conflict |
Cannot modify clock settings. The instrument is busy. |
|
-221 |
Settings Conflict |
Cannot modify input setting while DSIM channel is configured as output. |
|
-221 |
Settings Conflict |
Cannot modify output setting while DSIM channel is configured as input. |
|
-221 |
Settings Conflict |
Cannot modify trigger polarity when trigger source is not HW trigger or when the instrument is busy. |
|
-221 |
Settings Conflict |
Cannot modify trigger settings. The instrument is busy. |
|
-221 |
Settings Conflict |
DSIM: Cannot modify 'non runtime' setting during power calibration. |
|
-221 |
Settings Conflict |
DSIM: Cannot modify 'non runtime' setting while playing. |
|
-221 |
Settings Conflict |
DSIM: Cannot modify disabled setting. |
|
-221 |
Settings Conflict |
DSIM: Cannot modify or query settings. Instrument not configured or DSIM not enabled. |
|
-221 |
Settings Conflict |
DSIM: Cannot modify Parallel mode setting when in Serial mode. |
|
-221 |
Settings Conflict |
DSIM: Cannot modify read only setting. |
|
-221 |
Settings Conflict |
DSIM: Cannot modify reference frequency source. The setting is read only when clock source is set to external. |
|
-221 |
Settings Conflict |
DSIM: Cannot modify reference frequency. The setting is read only at this time. |
|
-221 |
Settings Conflict |
DSIM: Cannot modify Serial mode setting when in Parallel mode. |
|
-221 |
Settings Conflict |
DSIM: Cannot modify Word Alignment. The setting is disabled when word size is 16 bits or larger. |
|
-221 |
Settings Conflict |
Fading: Antenna settings are read only unless the correlation source is Custom Antenna Setup. |
|
-221 |
Settings Conflict |
Fading: Beam forming correlation type is only valid for 4x2 and 2x4 configurations. |
|
-221 |
Settings Conflict |
Fading: Cannot modify 'non runtime' setting while playing. |
|
-221 |
Settings Conflict |
Fading: Cannot modify disabled channel. |
|
-221 |
Settings Conflict |
Fading: Cannot modify disabled path. |
|
-221 |
Settings Conflict |
Fading: Cannot modify disabled setting. |
|
-221 |
Settings Conflict |
Fading: Cannot modify or query settings. Instrument not configured or Fading not enabled. |
|
-221 |
Settings Conflict |
Fading: Cannot modify read only setting. |
|
-221 |
Settings Conflict |
Fading: Correlation coefficients are read only unless the correlation source is Custom Correlation Matrix. |
|
-221 |
Settings Conflict |
Fading: Downlink direction is invalid for 2x4 configuration with Beam forming. |
|
-221 |
Settings Conflict |
Fading: High, Medium, and Low correlation types are only valid for 2x2 configuration. |
|
-221 |
Settings Conflict |
Fading: Only Laplacian Power Azimuth Spectrum is allowed for High, Medium, or Low correlation. |
|
-221 |
Settings Conflict |
Fading: Path Configuration settings are read only unless the path configuration source is Custom. |
|
-221 |
Settings Conflict |
Fading: The correlation type is read only unless the correlation source is set to Standard. |
|
-221 |
Settings Conflict |
Fading: The first channel index must be larger than the second channel index when setting a correlation coefficient. |
|
-221 |
Settings Conflict |
Fading: Uplink direction is invalid for 4x2 configuration with Beam forming. |
|
-221 |
Settings Conflict |
I/O: Cannot modify 'non runtime' setting during power calibration. |
|
-221 |
Settings Conflict |
I/O: Cannot modify 'non runtime' setting while playing. |
|
-221 |
Settings Conflict |
I/O: Cannot modify Analog Enabled when configured as a Digital Port. |
|
-221 |
Settings Conflict |
I/O: Cannot modify AWGN Settings. AWGN not licensed. |
|
-221 |
Settings Conflict |
I/O: Cannot modify input settings while I/O channel is configured as Output. |
|
-221 |
Settings Conflict |
I/O: Cannot modify LVDS Enabled when configured as an Analog Port. |
|
-221 |
Settings Conflict |
I/O: Cannot modify Marker 4 Settings when connected to a DSIM. The DSIM only uses Markers 1 and 3. |
|
-221 |
Settings Conflict |
I/O: Cannot modify or query settings. The instrument not configured or I/O not enabled. |
|
-221 |
Settings Conflict |
I/O: Cannot modify output settings while I/O channel is configured as Input. |
|
-221 |
Settings Conflict |
I/O: Cannot modify Power Meter settings in Auto Power Calibration mode. |
|
-221 |
Settings Conflict |
I/O: Cannot modify read only setting. |
|
-221 |
Settings Conflict |
I/O: Cannot modify setting. Marker 2 is reserved for internal use. |
|
-221 |
Settings Conflict |
Power meter: Setting disabled when Input RMS Source is not set to Measure. |
|
-221 |
Settings Conflict |
Power meter: Setting disabled when Input RMS Source is not set to User. |
|
-221 |
Settings Conflict |
Power meter: Setting disabled when Power Meter Gate is set to Threshold. |
|
-221 |
Settings Conflict |
Power meter: Setting disabled when Power Meter Gating is set to Marker. |
|
-221 |
Settings Conflict |
Power meter: Setting is read only when Input RMS Source is set to Waveform. |
|
-221 |
Settings Conflict |
Signal Analyzer: Cannot modify 'non runtime' setting during power calibration. |
|
-221 |
Settings Conflict |
Signal Analyzer: Cannot modify 'non runtime' setting while playing. |
|
-221 |
Settings Conflict |
Signal Analyzer: Cannot modify disabled setting. |
|
-221 |
Settings Conflict |
Signal Analyzer: Cannot modify or query settings. Instrument not configured or Signal Analyzer not enabled. |
|
-221 |
Settings Conflict |
Signal Analyzer: Cannot modify read only setting. |
|
-221 |
Settings Conflict |
Signal Generator: Cannot modify 'non runtime' setting during power calibration. |
|
-221 |
Settings Conflict |
Signal Generator: Cannot modify 'non runtime' setting while playing. |
|
-221 |
Settings Conflict |
Signal Generator: Cannot modify disabled setting. |
|
-221 |
Settings Conflict |
Signal Generator: Cannot modify or query settings. Instrument not configured or Signal Generator not enabled. |
|
-221 |
Settings Conflict |
Signal Generator: Cannot modify read only setting. |
|
-221 |
Settings Conflict |
The instrument is busy. |
|
-150 |
String Data Error |
Invalid master channel name: {0}. |
{0}=Master Channel Name |
1 |
Startup Error |
Interconnect board not found. |
|
1 |
Startup Error |
Invalid BB board arrangement. Arrange the boards to be contiguous beginning with slot 1 (back, next to clock board). |
Message is displayed when there is a non contiguous Baseband board in the PXB system. |
1 |
Startup Error |
Invalid I/O slot number: {0}. |
{0}=Slot number |
1 |
Startup Error |
One or more EEPROM boards not programmed correctly. If the problem persists, contact Agilent Technologies. |
|
1 |
Startup Error |
Power up self test failed. Run full self test for more details. |
|
1 |
Startup Error |
System hardware failure. Power down and restart the instrument. If the problem persists, contact Agilent Technologies. |
|
2 |
Shutdown Error |
An unexpected error has occurred. Power down and restart the instrument. If the problem persists, contact Agilent Technologies. |
|
3 |
Self Test Error |
Factory self test data not available. Power down and restart the instrument. If the problem persists, contact Agilent Technologies. |
|
3 |
Self Test Error |
Failed to run test {0} of test group {1}. Power down and restart the instrument. If the problem persists, contact Agilent Technologies. |
{0}=Test Name |
3 |
Self Test Error |
Failed to save test results. Power down and restart the instrument. If the problem persists, contact Agilent Technologies. |
|
4 |
Play Error |
Configuration license expired. |
|
4 |
Play Error |
Hardware optimization failed. |
|
4 |
Play Error |
Invalid configuration. |
|
4 |
Play Error |
Load Baseband Generator Setting Error. System hardware failure. Power down and restart the instrument. If the problem persists, contact Agilent Technologies. |
|
4 |
Play Error |
Load DSIM Setting Error. System hardware failure. Power down and restart the instrument. If the problem persists, contact Agilent Technologies. |
|
4 |
Play Error |
Load Fading Setting Error. Could not calculate the Cholesky decomposition for the correlation matrix. |
|
4 |
Play Error |
Load Fading Setting Error. Fading model not licensed. |
|
4 |
Play Error |
Load Fading Setting Error. Invalid dynamic fading file. |
|
4 |
Play Error |
Load Fading Setting Error. System hardware failure. Power down and restart the instrument. If the problem persists, contact Agilent Technologies. |
|
4 |
Play Error |
Load I/O Setting Error. System hardware failure. Power down and restart the instrument. If the problem persists, contact Agilent Technologies. |
|
4 |
Play Error |
Load Signal Analyzer Setting Error. External instrument {0} not connected or not powered on. |
{0}=Instrument Name |
4 |
Play Error |
Load Signal Generator Setting Error. External instrument {0} not connected or not powered on. |
{0}=Instrument Name |
4 |
Play Error |
Signal generator {0} did not arm properly. Check digital cable between PXB and signal generator {0}. |
{0}=Instrument Name |
4 |
Play Error |
Signal generator {0} did not arm in the allowed amount of time. |
{0}=Instrument Name |
4 |
Play Error |
The instrument not configured yet. |
|
5 |
Stop Error |
An unexpected error has occurred. Power down and restart the instrument. If the problem persists, contact Agilent Technologies. |
|
6 |
Save Error |
Could not open file {0}. |
{0}=File Name |
6 |
Save Error |
Save Baseband Generator Setting Error. Power down and restart the instrument. If the problem persists, contact Agilent Technologies. |
|
6 |
Save Error |
Save DSIM Setting Error. Power down and restart the instrument. If the problem persists, contact Agilent Technologies. |
|
6 |
Save Error |
Save Fading Setting Error. Power down and restart the instrument. If the problem persists, contact Agilent Technologies. |
|
6 |
Save Error |
Save I/O Setting Error. Power down and restart the instrument. If the problem persists, contact Agilent Technologies. |
|
6 |
Save Error |
Save Signal Analyzer Setting Error. Power down and restart the instrument. If the problem persists, contact Agilent Technologies. |
|
6 |
Save Error |
Save Signal Generator Setting Error. Power down and restart the instrument. If the problem persists, contact Agilent Technologies. |
|
7 |
Recall Error |
Could not open file {0}. |
{0}=File Name |
7 |
Recall Error |
Recall Baseband Generator Setting Error. Saved state corrupted or out of date. |
|
7 |
Recall Error |
Recall Configuration Error. Saved state corrupted or out of date. |
|
7 |
Recall Error |
Recall DSIM Setting Error. Saved state corrupted or out of date. |
|
7 |
Recall Error |
Recall Fading Setting Error. Could not calculate the Cholesky decomposition for the correlation matrix. |
|
7 |
Recall Error |
Recall Fading Setting Error. Fading model not licensed. |
|
7 |
Recall Error |
Recall Fading Setting Error. Invalid dynamic fading file. |
|
7 |
Recall Error |
Recall Fading Setting Error. Saved state corrupted or out of date. |
|
7 |
Recall Error |
Recall I/O Setting Error. Saved state corrupted or out of date. |
|
7 |
Recall Error |
Recall Signal Analyzer Setting Error. Saved state corrupted or out of date. |
|
7 |
Recall Error |
Recall Signal Generator Setting Error. Saved state corrupted or out of date. |
|
8 |
Copy Error |
Copy Baseband Generator Setting Error. |
|
8 |
Copy Error |
Copy DSIM Setting Error. |
|
8 |
Copy Error |
Copy Fading Setting Error. |
|
8 |
Copy Error |
Copy I/O Setting Error. |
|
8 |
Copy Error |
Copy Signal Analyzer Setting Error. |
|
8 |
Copy Error |
Copy Signal Generator Setting Error. |
|
9 |
Paste Error |
Paste Baseband Generator Setting Error. |
|
9 |
Paste Error |
Paste DSIM Setting Error. |
|
9 |
Paste Error |
Paste Fading Setting Error. |
|
9 |
Paste Error |
Paste I/O Setting Error. |
|
9 |
Paste Error |
Paste Signal Analyzer Setting Error. |
|
9 |
Paste Error |
Paste Signal Generator Setting Error. |
|
10 |
Reset Error |
An unexpected error has occurred. Power down and restart the instrument. If the problem persists, contact Agilent Technologies. |
|
20 |
Select Configuration Error |
Configuration corrupted. |
|
20 |
Select Configuration Error |
Configuration not found. |
|
20 |
Select Configuration Error |
Failed to retrieve installed licenses. |
|
20 |
Select Configuration Error |
Invalid configuration. |
|
21 |
Load Configuration Error |
Cannot apply config 'FADE (ext in) 1 Channel' because it has conflicting instrument assignments. |
Possible licensing problem (see Loopback Fading) |
21 |
Load Configuration Error |
Cannot use GPIB interface {0} as a system controller. It is not configured as a system controller. |
|
21 |
Load Configuration Error |
Configuration not licensed. |
|
21 |
Load Configuration Error |
DSIM not initialized properly. An unexpected error has occurred. Power down and restart the instrument. If the problem persists, contact Agilent Technologies. |
|
21 |
Load Configuration Error |
External instrument {0} and {1} have the same address ({2}). |
{0}=Instrument1 Name |
21 |
Load Configuration Error |
Failed to apply Fading model. Fading model {0} not licensed. |
{0}=Model Name |
21 |
Load Configuration Error |
Failed to apply Fading model. Failed to calculate the correlation coefficients. |
|
21 |
Load Configuration Error |
Failed to retrieve installed licenses. |
|
21 |
Load Configuration Error |
FPGA Alignment Failure |
|
22 |
Modify Configuration Error |
Analog input port not supported. |
|
22 |
Modify Configuration Error |
Configuration not found. |
|
22 |
Modify Configuration Error |
GPIB card {0} is configured as controller. |
{0} GPIB card number |
22 |
Modify Configuration Error |
GPIB card not found. |
|
22 |
Modify Configuration Error |
Instrument {0} not found. |
{0}=Instrument Name |
22 |
Modify Configuration Error |
Invalid analog output instrument type. Signal Generator or Custom instrument expected. |
|
22 |
Modify Configuration Error |
Invalid digital input instrument type. DSIM or Signal Analyzer expected. |
|
22 |
Modify Configuration Error |
Invalid digital output instrument type. DSIM or Signal Generator expected. |
|
22 |
Modify Configuration Error |
I/O port {0} not available. |
{0}=I/O Port Name |
22 |
Modify Configuration Error |
I/O port {0} not found. |
{0}=I/O Port Name |
30 |
Download Waveform Error |
Cannot start a download when another download is in progress. |
|
30 |
Download Waveform Error |
Could not open file {0}. |
{0}=Waveform File Name |
30 |
Download Waveform Error |
Failed to transfer waveform and marker data from files to dram. |
|
30 |
Download Waveform Error |
Marker file not found: {0}. |
{0}=Marker File Name |
30 |
Download Waveform Error |
The number of samples and the number of markers does not match. |
|
30 |
Download Waveform Error |
Waveform not found: {0}. |
{0}=Waveform File Name |
30 |
Download Waveform Error |
Waveform not licensed: {0}. |
{0}=Waveform Name |
31 |
Transfer Data Error |
Failed to transfer waveform data to instrument. Waveform name: {0}. |
{0}=Waveform Name |
40 |
Power Calibration Error |
Failed to abort power calibration. |
|
40 |
Power Calibration Error |
Failed to restore system state after a power calibration error. |
|
40 |
Power Calibration Error |
Instrument not configured yet. |
|
40 |
Power Calibration Error |
Power Calibration cannot complete, all inputs are disabled. |
|
40 |
Power Calibration Error |
Power Calibration setup failed for signal generator {0}. |
{0}=Instrument Name |
40 |
Power Calibration Error |
Power Meter Gating is set to Marker and the marker is turned off. |
|
40 |
Power Calibration Error, Power Search Error |
Power search failed for signal generator {0}. Increase I/Q signal level or reduce signal generator amplitude. |
{0}=Instrument Name |
40 |
Power Calibration Error |
Reading Power Meters timed out. |
|
40 |
Power Calibration Error |
The instrument is busy. |
|
40 |
Power Calibration Error |
Waveform RMS level exceeds signal generator {0} power search manual level maximum of {1}. |
{0}=Instrument Name |
40 |
Power Calibration Error |
Waveform RMS level is too low for successful signal generator {0} power search. |
{0}=Instrument Name |
41 |
I/O Calibration Error |
Failed to restore factory calibration data. An unexpected error has occurred. Power down and restart the instrument. If the problem persists, contact Agilent Technologies. |
|
41 |
I/O Calibration Error |
The instrument is busy. |
|
50 |
External Reference Input Error |
External reference frequency out of range. |
|
50 |
External Reference Input Error |
External reference source not detected. |
|
60 |
External Trigger Input Error |
Report this error to Agilent Technologies. |
|
100 |
System Error |
External reference source not detected. PLL unlocked. |
|
100 |
System Error |
Failed to configure GPIB address. GPIB card not found. |
|
100 |
System Error |
Failed to toggle touch screen on/off. Touch screen was not installed properly. |
|
100 |
System Error |
Invalid external reference frequency. PLL unlocked. |
|
100 |
System Error |
PLL unlocked. |
|
101 |
Hardware Error |
Clock board temperature {0} exceeds threshold. Shutting down system. |
{0}=Temperature |
101 |
Hardware Error |
Fans have stopped working. Shutting down system. |
|
101 |
Hardware Error |
I/O board temperature {0} exceeds threshold. Shutting down system. |
{0}=Temperature |
101 |
Hardware Error |
Mother board temperature {0} exceeds threshold. Shutting down system. |
{0}=Temperature |
101 |
Hardware Error |
Power supply temperature {0} exceeds threshold. Shutting down system. |
{0}=Temperature |
102 |
Add External Instrument Error |
Instrument {0} already exists. |
{0}=Instrument Name |
102 |
Add External Instrument Error |
Invalid instrument name. |
{0}=Instrument Name |
102 |
Add External Instrument Error |
Signal analyzer list is full. Remove an existing signal analyzer before adding a new one. |
|
102 |
Add External Instrument Error |
Signal generator list is full. Remove an existing signal generator before adding a new one. |
|
102 |
Add External Instrument Error |
The custom instrument list is full. Remove an existing custom instrument before adding a new one. |
|
102 |
Add External Instrument Error |
The DSIM list is full. Remove an existing DSIM before adding a new one. |
|
103 |
Modify External Instrument Error |
Instrument {0} not found. |
{0}=Instrument Name |
104 |
Remove External Instrument Error |
Instrument {0} in use. |
{0}=Instrument Name |
104 |
Remove External Instrument Error |
Instrument {0} not found. |
{0}=Instrument Name |
105 |
Test Connection Error |
Communication failure with instrument {0}. Instrument is busy, not connected or not powered on. |
{0}=Instrument Name |
110 |
Launch Application Error |
Failed to launch {0}. |
{0}=Application Name |
111 |
Launch License Manager Error |
Agilent License Server not installed properly. |
|
111 |
Launch License Manager Error |
Agilent License Server not installed. |
|
200 |
Warning |
{0} does not have external instrument yet. Please assign an instrument to this I/O port. |
{0}=I/O Port Name |
200 |
Warning |
Address {0} is already assigned to external instrument {1}. |
{0}=Instrument Address |
200 |
Warning |
Backup file corrupted or out of date. |
|
200 |
Warning |
Correlation magnitude was clipped to the maximum value of 1. |
|
200 |
Warning |
Correlation value must be a complex number with magnitude less than or equal to 1. |
|
200 |
Warning |
External instrument {0} and {1} have the same address ({2}). |
{0}=Instrument1 Name |
200 |
Warning |
Input Frequency Offset + Input Signal Bandwidth may exceed the effective fader bandwidth. Reduce the frequency offset or number of enabled fader paths. |
|
200 |
Warning |
Input Frequency Offset + Input Signal Bandwidth may exceed the output device bandwidth. |
|
200 |
Warning |
Input Frequency Offset + Output Frequency Offset may exceed the output device bandwidth. |
|
200 |
Warning |
Input Frequency Offset exceeds the effective fader bandwidth. Reduce the frequency offset or decrease the number of enabled fader paths. |
|
200 |
Warning |
Instrument {0} has invalid address. Correct at a later time. |
{0}=Instrument Name |
200 |
Warning |
Instrument {0} not connected or not powered on. |
{0}=Instrument Name |
200 |
Warning |
LXI LAN Reset Error |
|
200 |
Warning |
No hardware or license found. |
|
200 |
Warning |
No markers active. If you intend to use markers for ALC or RF blanking, you must turn them on. |
|
200 |
Warning |
Output Frequency Offset + Noise Bandwidth exceeds the output device bandwidth. |
|
200 |
Warning |
Power Meter Gating is set to Marker and the marker is turned off. |
|
200 |
Warning |
Recall external instruments failed. |
|
200 |
Warning |
The AWGN noise bandwidth exceeds the output device bandwidth. Reduce the noise bandwidth to {0} MHz. |
{0}=Output Device Bandwidth |
200 |
Warning |
The input frequency offset exceeds the output device bandwidth. Reduce the input frequency offset. |
|
200 |
Warning |
Touch screen not installed properly. |
|
200 |
Warning |
Waveform not licensed. Waveform name: {0}. |
{0}=Waveform Name |
201 |
Information |
{0} was clipped to {1} due to the valid resolution changing. |
{0}=Setting Name |
201 |
Information |
{0} was clipped to the maximum value of {1} |
{0}=Setting Name |
201 |
Information |
{0} was clipped to the minimum value of {1} |
{0}=Setting Name |
201 |
Information |
Baseband Generator: {0} clipped to {1}. |
{0}=Setting Name |
201 |
Information |
DSIM: {0} clipped to {1}. |
{0}=Setting Name |
201 |
Information |
I/O: Failed to copy master channel list. |
|
201 |
Information |
I/O: {0} clipped to {1}. |
{0}=Setting Name |
201 |
Information |
Power Calibration will run automatically on Play. |
|
201 |
Information |
Power control settings have been changed to match the current waveform type. |
|
201 |
Information |
Signal analyzer {0} is connected. |
{0}=Instrument Name |
201 |
Information |
Signal analyzer {0} is not connected. Address not found. |
{0}=Instrument Name |
201 |
Information |
Signal analyzer {0} is not connected. Address not valid. |
{0}=Instrument Name |
201 |
Information |
Signal analyzer {0} is not connected. Connection lost. |
{0}=Instrument Name |
201 |
Information |
Signal analyzer {0} is not connected. Model number not supported. |
{0}=Instrument Name |
201 |
Information |
Signal analyzer {0} is not connected. Not a member of the {0} family. |
{0}=Instrument Family |
201 |
Information |
Signal analyzer {0} is not connected. Requires firmware version {0} or higher and option (one of) {1}. |
{0}=Firmware version |
201 |
Information |
Signal analyzer {0} is not connected. Requires firmware version {0} or higher. |
{0}=Firmware version |
201 |
Information |
Signal analyzer: {0} clipped to {1}. |
{0}=Setting Name |
201 |
Information |
Signal generator {0} is connected. |
{0}=Instrument Name |
201 |
Information |
Signal generator {0} is not connected. Address not found. |
{0}=Instrument Name |
201 |
Information |
Signal generator {0} is not connected. Address not valid. |
{0}=Instrument Name |
201 |
Information |
Signal generator {0} is not connected. Connection lost. |
{0}=Instrument Name |
201 |
Information |
Signal generator {0} is not connected. Model number not supported. |
{0}=Instrument Name |
201 |
Information |
Signal generator {0} is not connected. Not a member of the {0} family. |
{0}=Instrument Family |
201 |
Information |
Signal generator {0} is not connected. Requires firmware version {0} or higher and option (one of) {1}. |
{0}=Firmware version |
201 |
Information |
Signal generator {0} is not connected. Requires firmware version {0} or higher. |
{0}=Firmware version |
201 |
Information |
Signal generator: {0} clipped to {1}. |
{0}=Setting Name |
201 |
Information |
Signal power calculations are invalid until a new power calibration is run. |
|
300 |
Modify Baseband Generator Setting Error |
Cannot modify waveform source created by Signal Studio. |
|
300 |
Modify Baseband Generator Setting Error |
I/P RMS Source invalid. Cannot extract I/P RMS information from binary waveforms. |
|
300 |
Modify Baseband Generator Setting Error |
Invalid marker file name: {0}. |
{0}=Marker File Name |
300 |
Modify Baseband Generator Setting Error |
Invalid waveform file name: {0}. |
{0}=Waveform File Name |
300 |
Modify Baseband Generator Setting Error |
Marker file not found: {0}. |
{0}=Marker File Name |
300 |
Modify Baseband Generator Setting Error |
Waveform file not found: {0}. |
{0}=Waveform File Name |
300 |
Modify Baseband Generator Setting Error |
Waveform size too small. It must have at least 256 samples. |
|
301 |
Load Baseband Generator Setting Error |
System hardware failure. |
|
302 |
Restore Baseband Generator Default Setting Error |
|
|
400 |
Modify Fading Setting Error |
Birth Death disabled when dynamic fading is enabled. |
|
400 |
Modify Fading Setting Error |
Dynamic Fading not allowed for faders using Birth Death. |
|
400 |
Modify Fading Setting Error |
Dynamic Fading not allowed for faders using High Speed Train. |
|
400 |
Modify Fading Setting Error |
Dynamic Fading not allowed for faders using Moving Propagation. |
|
400 |
Modify Fading Setting Error |
Fading model not licensed: {0}. |
{0}=Model Name |
400 |
Modify Fading Setting Error |
High bits of Random Seed must be between 0x0 and 0x1FFFFFF. |
|
400 |
Modify Fading Setting Error |
High Speed Train disabled when dynamic fading is enabled. |
|
400 |
Modify Fading Setting Error |
Low bits of Random Seed must be between 0x0 and 0xFFFFFFFF. |
|
400 |
Modify Fading Setting Error |
Middle bits of Random Seed must be between 0x0 and 0xFFFFFFFF. |
|
400 |
Modify Fading Setting Error |
Moving Propagation disabled when dynamic fading is enabled. |
|
401 |
Load Fading Setting Error |
Delay out of range. Expected [{0}, {1}]; found {2} |
{0} = minimum delay (currently 0 sec) |
401 |
Load Fading Setting Error |
Dwell Time out of range. Expected [{0}, {1}]; found {2} |
{0} = minimum dwell time (currently 0.01 sec) |
401 |
Load Fading Setting Error |
Fading Settings Download already in progress. |
|
401 |
Load Fading Setting Error |
Failed to calculate the Cholesky decomposition for the correlation matrix. |
|
401 |
Load Fading Setting Error |
Invalid dynamic fading file. |
|
401 |
Load Fading Setting Error |
Invalid dynamic fading file. Expected {0} parameters on line {1} ({2} columns x {3} paths), but {4} were found. |
{0} = expected number of parameters |
401 |
Load Fading Setting Error |
Invalid dynamic fading file. Number of dynamic fading states exceeds maximum of {0}. |
{0} = Maximum number of dynamic fading states allowed (currently 5000) |
401 |
Load Fading Setting Error |
Invalid Dynamic Fading File. The vehicle speed for each path must be the same unless the spectral shape is Jakes Rounded or Jakes Classical. |
|
401 |
Load Fading Setting Error |
Invalid Fader directive in dynamic fading file on line {0} - couldn't read number of paths. |
{0} = line number |
401 |
Load Fading Setting Error |
Loss out of range. Expected [{0}, {1}]; found {2} |
{0} = minimum loss (currently 0 dB) |
401 |
Load Fading Setting Error |
System hardware failure. |
|
401 |
Load Fading Setting Error |
UE Speed out of range. Expected [{0}, {1}]; found {2} |
{0} = minimum UE speed (currently 0 km/hr) |
401 |
Load Fading Setting Error |
Undefined parameter name in Columns directive of dynamic fading file on line {0}. |
{0} = line number |
402 |
Restore Fading Default Setting Error |
|
|
403 |
Apply Fading Model Error |
Fading model not licensed: {0}. |
{0}=Model Name |
403 |
Apply Fading Model Error |
Failed to calculate the correlation coefficients. |
|
500 |
Modify I/O Setting Error |
Cannot modify input setting while I/O channel is configured as output. |
|
500 |
Modify I/O Setting Error |
Cannot modify output setting while I/O channel is configured as input. |
|
500 |
Modify I/O Setting Error |
Failed to set {0} to {1} Setting invalid. Signal power calculations are invalid until a new power calibration is run. |
{0}=Setting Name |
501 |
Load I/O Setting Error |
System hardware failure. |
|
502 |
Restore I/O Default Setting Error |
|
|
601 |
Load Signal Generator Setting Error |
{0} not connected or not powered on. |
{0}=Instrument Name |
607 |
Query Signal Generator Setting Error |
{0} not connected or not powered on. |
{0}=Instrument Name |
608 |
Signal Generator Power Search Error |
{0} not connected or not powered on. |
{0}=Instrument Name |
609 |
Preset Signal Generator Error |
{0} not connected or not powered on. |
{0}=Instrument Name |
701 |
Load Signal Analyzer Setting Error |
{0} not connected or not powered on. |
{0}=Signal Analyzer Name |
702 |
Preset Signal Analyzer Error |
{0} not connected or not powered on. |
{0}=Signal Analyzer Name |
703 |
Query Signal Analyzer Setting Error |
{0} not connected or not powered on. |
{0}=Signal Analyzer Name |
801 |
Load DSIM Setting Error |
DSIM not connected or not powered on. |
|
801 |
Load DSIM Setting Error |
System hardware failure. |
|
802 |
Restore DSIM Default Setting Error |
DSIM not connected or not powered on. |
|
802 |
Restore DSIM Default Setting Error |
System hardware failure. |
|
These error messages can be displayed when you are completing the Microsoft Excel spreadsheet template for dynamic fading.
Message | Comment |
---|---|
Based on the carrier frequency of {0} and the maximum Doppler frequency of 1600 Hz, the UE Speed for Path {1} in row {2} must be between 0 and {3} km/hr. |
{0} = carrier frequency |
Delay for Path {0} in row {1} must be a numeric value. |
{0} = path number |
Delay for Path {0} in row {1} must be between 0 s and 2 ms. |
{0} = path number |
Dwell Time for row {0} must be a numeric value. |
{0} = row number |
Dwell Time for row {0} must be greater than or equal to 0.01 s. |
{0} = row number |
Error encountered during Export. Please correct the error and try again. |
|
Fader {0} Carrier Frequency must be a numeric value. |
{0} = fader number |
Loss for Path {0} in row {1} must be a numeric value. |
{0} = path number |
Loss for Path {0} in row {1} must be between 0 and 84 dB. |
{0} =
path number |
The file you have specified for export is marked as read only. |
|
The maximum number of dynamic fading states 5000, but fader {0} has {1} states. Exporting the first 5000 states. |
{0} = fader number |
UE Speed for Path {0} in row {1} must be a numeric value. |
{0} = path number |